High-purity water reference sheets.

The figures used most often in the field, one page each. Limits are drawn from the same standards source Eighteen2 checks designs against. Print them for the mechanical room.

Seven one-page references for the people who walk the mechanical room. Each prints on its own page.

Sterile processing

AAMI ST108 water quality

ST108 defines two water qualities for device reprocessing. Critical water is for the final rinse and anywhere water touches a device after cleaning. Utility water is for washing, flushing, and intermediate rinses.

Critical water
ParameterLimit
pH @ 25 °C5.0 – 7.5
Conductivity< 10 µS/cm
Total alkalinity< 8 mg/L as CaCO₃
Total hardness< 1 mg/L as CaCO₃
Bacteria< 10 CFU/mL
Endotoxin< 10 EU/mL
TOC< 1.0 mg/L
Chloride< 1 mg/L
Iron, copper, manganese, zinc, aluminum< 0.1 mg/L each
Nitrate, phosphate, sulfate, silicate< 1 mg/L each
Color and turbidityColorless, clear, no sediment
Utility water (washing, flushing, intermediate rinse)
ParameterLimit
pH @ 25 °C6.5 – 9.5
Conductivity< 500 µS/cm
Total alkalinity< 400 mg/L as CaCO₃
Total hardness< 150 mg/L as CaCO₃Soften above this unless detergent chemistry tolerates it.
Bacteria< 500 CFU/mL< 10 CFU/mL and < 10 EU/mL when used as final rinse after chemical HLD.
Chloride< 250 mg/L
Silicate< 50 mg/L

Steam

ST108 also addresses the water used to make steam for sterilization. Size the steam generator's feed to the sterilizer maker's instructions for use and the facility's water management plan, and record which source feeds which sterilizer.

Monitoring at a glance

  • Conductivity · Daily · After last treatment step
  • pH, alkalinity, hardness · Monthly · After last treatment step
  • Bacteria (HPC) and endotoxin (LAL) · Monthly · Loop out and loop return
  • pH, conductivity, alkalinity, hardness; bacteria and endotoxin · Monthly · First point of use (chemistry); each point of use (bacteria, endotoxin)
  • Utility water chemistry and bacteria · Quarterly · Treatment system and point of use

Frequencies are the standard's recommended minimums. A water management team and a documented risk assessment are expected.

AAMI ST108 (ST108:2023), Water for the processing of medical devices. Limits as summarized in the Eighteen2 standards source.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

Laboratory

ASTM D1193 reagent water, Types I–III

Types I, II, and III are defined by ionic, organic, and silica limits. Bacterial subtypes A, B, and C are optional add-ons; ask whether the customer's spec names one.

ParameterType IType IIType III
Resistivity @ 25 °C≥ 18.0 MΩ·cm≥ 1.0 MΩ·cm≥ 4.0 MΩ·cm
Conductivity @ 25 °C≤ 0.056 µS/cm≤ 1.0 µS/cm≤ 0.25 µS/cm
TOC≤ 50 µg/L≤ 50 µg/L≤ 200 µg/L
Sodium≤ 1 µg/L≤ 5 µg/L≤ 10 µg/L
Chloride≤ 1 µg/L≤ 5 µg/L≤ 10 µg/L
Silica≤ 3 µg/L≤ 3 µg/L≤ 500 µg/L
Final filtration0.2 µmNot specified0.45 µm

Rules of thumb

  • Type I is usually made at the point of use: an RO or Type II feed, then a mixed-bed polish, UV, and a 0.2 µm final filter.
  • Type II: older editions called for distillation; current editions accept any process that meets the limits. Confirm the customer spec does not still require distilled water.
  • Type III at 4 MΩ·cm is beyond RO alone. Plan a DI polish.
ASTM D1193 (D1193-06 (2018)), Standard Specification for Reagent Water. D1193-24 is the current edition; limits shown follow the 2006 table.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

Clinical lab · Pharma

CLSI GP40 and USP Purified Water

CLSI clinical laboratory reagent water (CLRW)
ParameterLimit
Resistivity @ 25 °C> 10 MΩ·cm
TOC< 500 ppbOnline TOC not required; referral-lab testing acceptable.
Bacteria< 10 CFU/mLViable plate count; endotoxin optional.
Particulates0.2 µm filtration or finer

Special reagent water (SRW) is set per test by the method manufacturer. Analyzer feed specs can be tighter than CLRW; carry both.

USP Purified Water
ParameterLimit
Conductivity (Stage 1 @ 25 °C)≤ 1.3 µS/cm
TOC≤ 500 ppb
Bacteria (action level)100 CFU/mL

Conductivity is a three-stage test; Stage 1 is read without temperature compensation, and 1.3 µS/cm applies at 25 °C. Purified Water systems need design qualification and IQ/OQ/PQ.

CAP inspections

CAP inspects against CLSI GP40 or the analyzer maker's spec. The lab defines the water type for each test, reviews resistivity daily, keeps bacterial counts on schedule, and documents every out-of-spec result and its corrective action. Confirm item numbers against the lab's current Laboratory General checklist.

CLSI GP40 (GP40-Ed5 (Dec 2024)); USP–NF <645> Water Conductivity and <643> Total Organic Carbon. CAP checklist items vary by edition.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

Technology

RO, DI, or CEDI

Reverse osmosisExchange DICEDI
What it removesMost dissolved salts, plus organics, particles, bacteria, and endotoxinIons, down to very high resistivityIons remaining after RO, continuously
Typical rolePrimary treatment ahead of storageLow flows, polishing, or labs without drain capacityChemical-free polish after RO
Needs ahead of itSediment filtration, softening or antiscalant, and chlorine removalIdeally RO; consumption scales with feed TDSRO permeate within the maker's hardness, CO₂, and chlorine limits
DrainYes, continuous reject; recovery set by feed chemistryNone on siteSmall concentrate stream
Watch forScaling, chlorine damage to thin-film membranes, cold-feed derateExhaustion between exchanges, resin fines, bacterial growth; add a post filterHardness and CO₂ breakthrough from upstream

Single-pass RO permeate is rarely below a few µS/cm. If the spec needs MΩ·cm, plan a polish.

General engineering practice. Confirm feed limits and performance with each manufacturer's projection.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

System design

A typical pretreatment and distribution system

  1. Backflow preventer (RPZ)Required by plumbing code where treatment connects to potable water.
  2. Pressure-reducing valveHolds a steady inlet for the equipment downstream.
  3. Sediment filterCommonly 5 µm. Protects softener valves and carbon beds.
  4. SoftenerTwin-tank for continuous service. Removes hardness that would scale the RO.
  5. CarbonRemoves chlorine. Chloramine needs catalytic carbon and more contact time; size on empty-bed contact time.
  6. ROPump, membranes, and a reject line to drain.
  7. Storage tankWith a hydrophobic vent filter, sized to cover peak demand.
  8. Distribution pumpOften duty/standby. Keeps the distribution loop moving.
  9. PolishDI or CEDI when the grade needs it.
  10. UV254 nm for bacterial control; 185 nm when TOC reduction is the goal.
  11. Final filter0.2 µm where the standard calls for it.
  12. Loop and points of useReturn to the tank, with sample valves at the loop out, the return, and each point of use.

Before you select anything, get a feed-water analysis: hardness, alkalinity, iron, silica, TDS, chlorine and chloramine, temperature range, and pressure.

Typical arrangement for commercial and healthcare RO systems. The feed-water analysis sets the final system.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

Units

Conductivity, resistivity, and hardness

Resistivity (MΩ·cm) = 1 ÷ conductivity (µS/cm). Use conductivity for treated and utility water, and resistivity once the water is high-purity.

ResistivityConductivityWhere it shows up
18.2 MΩ·cm0.055 µS/cmTheoretical pure water at 25 °C
18.0 MΩ·cm0.056 µS/cmASTM Type I floor
10 MΩ·cm0.1 µS/cmCLSI CLRW floor
4 MΩ·cm0.25 µS/cmASTM Type III floor
1 MΩ·cm1 µS/cmASTM Type II floor
0.77 MΩ·cm1.3 µS/cmUSP <645> Stage 1 limit at 25 °C
0.1 MΩ·cm (100 kΩ·cm)10 µS/cmAAMI ST108 critical water limit
0.002 MΩ·cm (2 kΩ·cm)500 µS/cmAAMI ST108 utility water limit

Handy conversions

  • Hardness: 1 grain per gallon (gpg) = 17.1 mg/L as CaCO₃.
  • TDS estimate: mg/L ≈ 0.5 to 0.7 × conductivity in µS/cm, depending on the ions present. Use a lab analysis for design.
  • Temperature: conductivity rises roughly 2% per °C in ordinary water, and faster in high-purity water. Always compare compensated readings.
  • ppm and ppb: in dilute water, 1 ppm ≈ 1 mg/L and 1 ppb ≈ 1 µg/L.
Unit definitions. Read conductivity and resistivity temperature-compensated to 25 °C, except USP <645> Stage 1, which is read uncompensated.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

Sterile processing

SPD equipment: what to ask

For each machine the system will feed, get these before you size anything.

  • Make and model, and the water spec in its instructions for use
  • Water quality per connection: utility, critical, or softened; hot or cold
  • Gallons per cycle and cycles per day, at today's load and planned growth
  • Peak flow (gpm) and how many machines can fill at once
  • Inlet pressure range and connection size
  • Inlet temperature, and whether the machine heats its own water
  • Drain capacity and location for the treatment equipment
  • Power available at the equipment room

Equipment that usually needs water

  • Washer-disinfectors
  • Steam sterilizers and steam generators
  • Cart and rack washers
  • Ultrasonic cleaners
  • Automated endoscope reprocessors
  • Manual cleaning sinks and rinse stations
Discovery checklist. The equipment maker's instructions for use set the water each machine needs.eighteen2.com · Reference summary, not a substitute for the standard. Designs are budgetary.

These sheets are working summaries for reps and engineers. They are not a substitute for the current edition of each standard or for the equipment maker's instructions for use, and they are not legal or regulatory advice. Eighteen2 designs are budgetary and are not stamped; a licensed engineer or the manufacturer confirms the final design.